The EC1000S not only supplies AC and DC power as the ordinary power supply, it also allows free programming of outputs such as instantaneous interruption, voltage sweep, and voltage variation patterns. The EC1000S also has essential functions for power tests, including a variety of output measurements and measurements related to the load power supply input. In addition, while the EC1000S can output as much as 1kVA*, it is desktop-size small and light, and affordable.
This programmable AC/DC power supply from NF Corporation can be used for a broad range of applications, while achieving excellent cost performance.
*The power supply outputs 1 kVA when the input is AC 200 V and 750 VA when the input is AC 100 V
Rich Measurement Functions
The EC1000S is equipped with high accuracy measurement functions for measuring frequencies, load power factors, crest factors (CF), and even the harmonic currents, in addition to the voltage and current. Setting values and measurement values are simultaneously displayed on easy-to-see 5.7-inch large-screen display.
Measures
Measurement Results and Setting Values | Measurement of Output Harmonic Current |
Freely program output patterns with the sequece function
Program, in advance in the main unit, output variation patterns used to test the power supplies of devices and parts (one sequence each for the AC mode and DC mode). Simultaneous sweeping of frequency and voltage, and arbitrary waveforms are also supported. The same tests can be reproduced regardless of time or operator.
Overview of seauence function
| Setting screen for the sequence function |
|
Example of voltage fluctuation test pattern
Control software allows smoother and more sophisticated customization of tests and measurements
The provided control software has (1) a logging function for importing and saving measurement value data, facilitating creation of reports, data analysis and other operations; (2) a sequence function to freely edit, execute, save and otherwise operate a series of output variation patterns; and (3) an arbitrary waveform creation function to easily create complicated waveforms. This software readily supports more sophisticated tests and measurements during development and prototyping
Remote Control | Logging of measurement values |
Sequence editing | Creation of arbitrary waveform |
A variety of output modes
Eight output modes combining two operation modes and four signal source modes can be set. The most appropriate mode for an application can be selected.
Operation modes | Alternating current (AC), direct current (AC + DC) |
---|---|
Signal source modes | Internal (INT), external (EXT), internal/external (ADD), external synchronization (SYNC) |
A variety of tests and measurements for electronics with a wide range of AC output voltages and output frequencies can be performed.
AC output voltage range | Maximum output current |
---|---|
100V range: 0V to 135V*, resolution of 0.1V | 10 A* |
200V range: 0V to 270V*, resolution of 0.1V | 5 A* |
*Power input AC 200V
Output frequency range (internal, internal/external) | Resolution |
---|---|
1 Hz to 550 Hz | 0.1 Hz |
In the DC mode, the EC1000S can output power which superimposes an alternating current (sine, square, or arbitrary wave) over a direct current. It can also be used as a DC power supply. This mode is best suited to DC/DC converter tests and ripple tests for capacitors.
DC output voltage range | Maximum output current |
---|---|
100V range: -190V to 190V*, resolution of 0.1V | 10 A |
200V range: -380V to 380V*, resolution of 0.1V | 5 A |
*Power Input AC 200 V
A peak current equal to four times the maximum output current to the capacitor input rectifier load (crest factor: 4) can be supplied.
The output current can be limited by setting the maximum output current (peak/average current) in advance. Prototypes can be protected from abnormal currents during development evaluation.