구입 및 기술문의 : 010-4221-0660
기본확도:0.02% o full scale
측정대역:DC, 0.5Hz~300kHz
10측정 데이터/100s의 고속통신·응답
1000Vrms의 고전압 측정
파형 출력 기능 탑재(option)
모터 평가 기능 탑재(WT1030M)
고조파 분석기능 :10~440Hz(option)
정격치(range)
전압:15/30/60/100/150/300/600/1000V
전류:직접입력:0.5/1/2/5/10/20A
전류:외부입력(option):250m/500m1/2.5/5/10V
주파수 범위
DC및 0.5Hz~300kHz(단, 전력은 200kHz까지)
기본확도(45Hz≤f≤66Hz)
전압/전류:±(0.1% of rdg+0.1% of rng)
전력:±(0.1% of rdg+0.1% of rng)
역률의 영향(cosφ=0일때):±0.15% of rdg가산
외형크기:426(W)×132(H)×400(D)mm
질량:약 10kg(3 element model)
질량:약 9kg(1 element model)
Figure 1 A model in the series of WT1000 |
With the increased capacity of power devices to handle wattage and frequency, and with the advanced techniques developed to control these devices, modern motor-controlling inverters have been designed to consume less energy, perform at higher levels and have enhanced functionality. Because of such reasons as the need to conserve energy, an increase in carrier frequency, and motor control at lower speeds by means of advanced control techniques, power meters must be capable of measuring electrical energy with high accuracy, handling broader bandwidths, and measuring power at lower frequencies.
The recently developed WT1000 series of digital power meters (models WT1010, WT1030 and WT1030M) has a measuring accuracy and bandwidth high and wide enough to measure the power consumption of inverter-driven motors, as well as a generous functionality that can be applied to a wide range of uses. Figure 1 is an external view of a model in the series of WT1000 digital power meters.
WT1000 digital power meters have the following main features:
Figure 2 Block Diagram of WT1000 Series of Digital Power Meters
Figure 2 is a basic block diagram of the WT1000 series of digital power meters. The entire circuitry consists of such building blocks as the input, DSP, CPU, display and power supply sections. In the input section, an input signal is converted to a digital value. Then, at the DSP, corrective calculation, square-law calculation, multiplication and addition are performed for each sampling interval to determine the rms values of voltage and current and the active power. When the power meter enters a display update interval, it calculates such parameters as voltage, current, and active, apparent and reactive powers, from the data resulting from the calculations noted earlier. Thus, the CPU processes these data to send to the display, communication line, printer and D/A output.
Figure 3 Power vs. Frequency Characteristics (Power Factor = 1) | Figure 4 Power vs. Frequency Characteristics (Power Factor = 0) |
Figure 5 Measuring Principle |
1. Input Sections
An input section consists of voltage- and current-input circuits, each isolated from each other. The voltage-input circuit employs a scheme of resistance-based potential division, while the current-input circuit uses a shunt-based scheme. Input to each circuit is normalized by an operational amplifier to turn it into digital values through a 16-bit A/D converter. The signals of these digital values are then isolated by the photo-isolator to be supplied to the DSP section. The A/D converter converts the input at a rate of approximately 60 kHz.
The input resistance of the voltage-input circuit is 2.4 MΩ, which comprises 24 units of 100-kΩresistors. This configuration is employed in order to keep the voltage applied to each of these resistors low so that less expensive resistors can be used with this circuit. The shunt of the current-input circuit has a resistance of 5-mΩ, where two 10-mΩresistors are connected in parallel so that currents with polarities opposite each other flow through these resistors. This configuration is employed in order to reduce the effects of a magnetic field formed by a large, high-frequency current, thus providing excellent frequency and phase characteristics. Figures 3 and 4 show the power vs. frequency characteristics.
2. DSP Section
Data converted to digital values at the input section undergo a corrective calculation against input-circuit errors, and the calculations of active power and the rms values of voltage and current, and then the calculations of apparent and reactive powers, power factor, and so on, at the DSP section. The following are the formulas for these calculations:
where v (k) = instantaneous value of voltage in k th sampling interval i (k) = instantaneous value of current in k th sampling interval N = number of sampling cycles over an integral multiple of the period of the input signal.
The active power and rms values are calculated using data sampled over an integral multiple of the period of the input signal. This interval is determined by the zero-cross signal derived from the input signal. If the input signal frequency becomes more than half as high as the sampling frequency, the period of the input signal fails to match that of the waveform reproduced after sampling, resulting in an error in the measured value. To solve this problem, the zero-cross signal of the input signal is synchronized with the sampling clock of the A/D converter to produce another zero-cross signal for the waveform being reproduced. The interval over which calculations are performed is thus determined from the period of this zero-cross signal. Figure 5 shows the measuring principle represented by the time relationships.
Figure 6 Example of WT1030M Digital Power Meter Wired for Motor-characterizing Capability
1. Motor-characterizing Capability
Traditionally, power consumption, torque and rotational speed have been measured separately in the characterization testing of motors. In this method, the time at which torque and rotational speed were measured was not consistent with that at which power was measured, requiring adjustments between these times.
The motor-characterizing capability of the new WT1030M digital power meter measures both power and torque almost simultaneously if the torquemeter's response is fast enough, compared with 100 ms, eliminating the need for such adjustments. The power meter can calculate the output of motors from the torque and rotational speed to figure out the efficiency between the input and output of a motor, or the efficiency between the inputs and outputs of an inverter and its inverter- driven motor system in the case of a single-phase input inverter. The meter can also calculate synchronous speed and slippage, enabling users to obtain all required data only from their power meter. Using the motor-characterizing capability, users can test motors with ease to characterize them. Figure 6 is an example of the WT1030M digital power meter wired for its motor- characterizing capability.
2. Harmonics Analysis
Along with an increasing concern over harmonic current measurements, the latest sampling power meters have been provided with a harmonics-analyzing capability. The WT1000 series discussed in this paper also has this capability. The harmonics-analyzing capability of these power meters can measure not only the conventional harmonics components, harmonics content and distortion factor but also phase angles formed between elements, as well as the apparent and reactive powers and the power factor of a fundamental wave which are difficult to measure in normal measurement. The measurable range of fundamental frequencies has been extended to a low of 10 Hz, compared with the conventional limit of 40 Hz. This improvement has made it possible to use these power meters not only in harmonics current measurements of commercial power sources but also in other areas of application. In addition, the D/A output can provide signals of harmonic components, distortion factors and harmonics content, permitting logging of variations over a prolonged period of time using a recorder.
등록된 상품이 없습니다.